Invacare Essential Basic Profiling Mattress User Manual

Invacare Essential mattress series overview

The Invacare Essential mattresses are designed to provide effective pressure redistribution for adults and adolescents with restricted or limited mobility. The mattresses feature a multi-stretch, water-resistant cover and a foam core to help promote patient comfort and reduce tissue pressure at vulnerable areas.

Özellikler

Amaçlanan kullanım Effective pressure redistribution as part of an overall pressure injury prevention program; used with an appropriately sized bed frame.
Hedeflenen kullanıcılar Adults and adolescents with restricted or limited mobility and/or activity (hospital, long-term care, or homecare).
Model variants / indications
  • Temel Temel: suitable for users who do not require therapy for pressure injury.
  • Essential Care, Visco, Touch: suitable for users “At Risk/High Risk” of developing pressure injury.
Service life (expected) Five years when used daily and in accordance with safety/maintenance/correct use instructions (effective service life may vary with frequency and intensity of use).
Cover and access Zipped around 3 sides for ease of inspection and cleaning; water-resistant, vapour-permeable, multi-stretch surface.
Essential Touch feature Double-sided castellated foam allows turning/rotation inside the cover for increased product longevity.
Safety/medical device marking CE mark (Medical Device Regulation 2017/745, Class I) and UKCA mark (UK MDR 2002 (as amended), Class I).
Malzemeler
  • Foam: fire-modified polyurethane foam
  • Cover: polyester fabric with polyurethane coating
  • Latex: none of the product components contains natural rubber latex
General size availability Mattresses are available in various sizes depending on the market; contact an Invacare distributor for detailed information.
Dimensions (from listed technical data ranges)
  • Uzunluk: 1780, 1950, 2000, 2050, and 2200 mm (varies by model/size)
  • Genişlik: 780, 880, 830, 850, 900, 1030, 1180 mm (varies by model/size)
  • Yükseklik: 120 or 150 mm (varies by model/size)
Maximum user weight (from listed technical data ranges) Up to 150 kg (values vary by model/size; multiple ranges are listed in the technical data table).
Product weight (from listed technical data ranges) Approx. 8 to 15 kg (values vary by model/size; multiple ranges are listed in the technical data table).

Quick setup and use

  1. Kullanım kılavuzunu okuyun and ensure you understand the warnings and instructions before use.
  2. Kullanıma hazırlanın: tüm ambalajları çıkarın.
  3. Place the mattress directly on the bed mattress support, or place the mattress overlay on the existing mattress.
  4. Allow rolled products to recover before use; follow the packaging insert instructions.
  5. Ensure correct orientation: make sure the printed side of the mattress cover faces upwards.
  6. Maintain required clearance: ensure the distance between the surface of the mattress and the top of the side rail is at least 220 mm.
  7. Keep the cover on: the mattress must be used with the cover at all times.
  8. Reposition regularly: insufficient repositioning can lead to tissue compression and potential injury formation; the patient (or caregiver) must reposition on a regular basis.
Loading PDF…
Download PDF

Sorun giderme

  • If the product shows defects: before use, check for defects; do not use a defective product.
  • If there is fluid ingress, stains, rips, or damage (“strike-through”): after each patient use and at least monthly (whichever occurs first), inspect foam and cover. Replace any damaged cover or stained foam.
  • If the cover is damaged with holes: there is a risk liquids may ingress and contaminate the mattress—ensure the mattress is not jammed or damaged by sharp edges and do not use sharp instruments/objects on or under the mattress.
  • After contamination: contact a hygiene specialist; remove contaminated foams from use.
  • If disinfection may have failed: ensure all cleaning agents and disinfectants are thoroughly rinsed off and dry thoroughly before use, since failure in disinfection can accumulate reagents that may damage the polyurethane coating or negate biocompatibility results.
  • If you find damage or malfunction: do not reuse the product and follow the replacement/reconditioning steps only when appropriate.

SSS

Must the mattress be used with the cover on?
Yes. The manual states that the mattress must be used with the cover at all times.
What should I do before using the mattress for the first time?
Remove all packaging, check the product for defects, ensure the printed side of the cover faces upwards, place it correctly on the bed support (or as an overlay), allow rolled products to recover, and ensure the distance to the side rail is at least 220 mm.
How often should the mattress be inspected and how?
Check for strike-through (e.g., fluid ingress, stains, rips, or damage) after each patient and on a minimum monthly basis (depending on which occurs first) by a suitably qualified and competent person. For overlays, unzip the cover completely and inspect staining on the white underside and interior foam.
How is the Essential Touch mattress different?
Essential Touch has double-sided castellated foam that allows the mattress to be turned or rotated inside the cover for increased product longevity.
Can the mattress be reused for another user?
Yes. It is suitable for reuse, but it must be reconditioned for a new user by performing inspection per the service plan and cleaning and disinfection as described in the maintenance section. If any damage or malfunction is detected, do not reuse.

Belgeler / Kaynaklar

PDF thumbnailEssential Basic Profiling Mattress
User Manual · DTEC013369, 1633734H, Essential Basic Profiling Mattress, Basic Profiling Mattress, Profiling Mattress, Mattress

Referanslar

Bir Soru Sorun

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

Bir Soru Sorun

Ask about setup, compatibility, troubleshooting, or anything missing from this manual. Name and email are optional.