VWR 89202-918

VWR Kuzey Amerika 89202-918 Sirkülasyonlu Açık Banyo Sıcaklık Kontrol Cihazı Kullanım Kılavuzu

Model: 89202-918

1. Giriş

This manual provides essential instructions for the safe and efficient operation, installation, and maintenance of the VWR North America 89202-918 Circulating Open Bath Temperature Controller. Please read this manual thoroughly before using the device to ensure proper function and to prevent potential hazards.

2. Güvenlik Bilgileri

Her zaman aşağıdaki güvenlik önlemlerine uyun:

3. Ürün Bittiview

The VWR 89202-918 is a precision temperature controller designed for use with open baths, providing accurate and stable temperature regulation for various laboratory applications. It features an intuitive control panel and a robust heating and circulation system.

VWR 89202-918 Circulating Open Bath Temperature Controller main unit

Şekil 1: Ön view of the VWR 89202-918 Circulating Open Bath Temperature Controller. This image displays the main control unit with its digital display and control buttons, along with the power cord extending from the base.

Heating element and circulator pump of VWR 89202-918

Şekil 2: Alt view of the VWR 89202-918 showing the heating coil and circulation components. This image highlights the internal heating element and the pump mechanism responsible for fluid circulation within the bath.

Product label for VWR 89202-918 showing model and electrical specifications

Şekil 3: Product label detailing the VWR North American Cat. No. 89202-918, model MX20S135-V11B, serial number, and electrical specifications (60 Hz/A/V, 120V). This label provides critical identification and power requirements for the device.

3.1. Bileşenler

4. Kurulum

  1. Paketin açılması: Carefully remove the controller from its packaging. Inspect for any visible damage.
  2. Montaj: Securely attach the controller to the side of your open bath using the integrated mounting clamp. Ensure the heating element and pump are fully submerged in the bath fluid, but the control head remains above the fluid level.
  3. Sıvı Seviyesi: Fill the open bath with the appropriate heat transfer fluid to a level that covers the heating element and pump, but is below the minimum fill line indicated on the controller (if applicable) and well below the control head.
  4. Güç Bağlantısı: Connect the power cord to a grounded electrical outlet matching the voltage specifications on the product label (120V, 60Hz).

5. Kullanım Talimatları

5.1. Güç Açma/Kapatma

5.2. Sıcaklığın Ayarlanması

  1. With the unit powered on, the display typically shows the current bath temperature.
  2. Basın AYARLAMAK button (or equivalent) to enter temperature setting mode. The setpoint temperature will flash.
  3. Kullanın UP Ve AŞAĞI İstenilen sıcaklık ayar noktasını ayarlamak için ok tuşlarını kullanın.
  4. Basmak AYARLAMAK again to confirm the new setpoint. The unit will begin heating or cooling to reach the target temperature.

5.3. Circulation Control

The circulation pump typically operates continuously when the unit is powered on to ensure uniform temperature. Refer to the specific model's advanced settings for pump speed adjustments if available.

6. Bakım

7. Sorun Giderme

SorunOlası NedenÇözüm
Ünite açılmıyor.No power supply; damaged power cord; internal fuse.Check power connection and outlet. Inspect power cord. Contact service if fuse replacement is suspected.
Sıcaklık sabit değil veya yanlış.Insufficient fluid level; contaminated fluid; sensor issue; external drafts.Ensure proper fluid level. Replace fluid if contaminated. Minimize external temperature fluctuations. Contact service if sensor is faulty.
Circulation pump not working.Blocked impeller; motor failure.Disconnect power and inspect impeller for obstructions. Contact service if motor failure is suspected.
Hata kodu görüntülendi.Belirli sistem hatası.Note the error code and refer to the full service manual or contact VWR technical support for specific guidance.

8. Özellikler

9. Garanti ve Destek

For warranty information, technical support, or service inquiries, please contact VWR North America directly or refer to the official VWR website. Keep your purchase receipt and product serial number (found on the product label, see Figure 3) readily available when contacting support.

VWR North America Contact: us.vwr.com/store/

İlgili Belgeler - 89202-918

Önview Gelişmiş Programlanabilir Sıcaklık Kontrollü VWR Signature Sirkülasyonlu Banyolar Kullanım Kılavuzu
Bu kullanım kılavuzu, hassas laboratuvar sıcaklığı kontrolü için tasarlanmış Gelişmiş Programlanabilir Sıcaklık Kontrol Cihazlarına sahip VWR Signature Sirkülasyon Banyolarının kurulumu, çalıştırılması ve bakımı için kapsamlı rehberlik sağlar.
Önview VWR MX Daldırma Sirkülatörü Kullanım Kılavuzu
VWR MX Daldırma Sirkülatörü için laboratuvar banyolarında hassas sıcaklık kontrolü için kurulum, çalıştırma, bakım, sorun giderme ve teknik özelliklerin ayrıntılı olarak açıklandığı kullanım kılavuzu.
Önview VWR Signature Circulating Baths with Advanced Digital Temperature Controller Instruction Manual
This instruction manual provides detailed guidance on the installation, operation, safety, and maintenance of VWR Signature Circulating Baths equipped with an Advanced Digital Temperature Controller. Learn about model specifications, troubleshooting, and technical information for precise temperature control applications.
Önview VWR 12 ve 18 Litre Lineer Çalkalamalı Su Banyosu Kullanım Kılavuzu
VWR 12 ve 18 litrelik doğrusal çalkalama su banyoları için kapsamlı kullanım kılavuzu. Güvenlik önlemleri, kurulum, çalıştırma, teknik özellikler, bakım, sorun giderme, garanti ve bertaraf konularını kapsar.
Önview VWR Genel Amaçlı Su Banyoları Kullanım Kılavuzu | Laboratuvar Sıcaklık Kontrolü
VWR Genel Amaçlı Su Banyoları için kapsamlı kullanım kılavuzu; laboratuvar inkübasyon, inaktivasyon ve aglütinasyon işlemleri için kurulum, çalıştırma, bakım, sorun giderme ve güvenlik yönergelerini kapsamaktadır.
Önview VWR Genel Amaçlı Su Banyoları Kullanım Kılavuzu
Operator's manual for VWR General Purpose Water Baths, detailing installation, operation, safety, maintenance, and troubleshooting for laboratory use.