Extech FL700

Extech FL700 Florür Ölçüm Cihazı Kullanım Kılavuzu

Modeli: FL700

1. Giriş

The Extech FL700 Fluoride Meter is a precision instrument designed for the rapid and accurate measurement of fluoride ions in various aqueous samples, including drinking water. This device integrates a sensing electrode, measurement electronics, and a digital display into a single, portable unit, making it suitable for both field and laboratory applications.

The FL700 provides direct readings of total fluoride concentrations ranging from 9.99 ppm or mg/L down to 0.1 ppm. It features automatic temperature compensation to ensure measurement accuracy across varying environmental conditions.

2. Güvenlik Bilgileri

Please read and understand all safety precautions before operating the Extech FL700 Fluoride Meter. Failure to follow these instructions may result in injury or damage to the instrument.

  • Always handle the fluoride electrode with care. Avoid touching the sensing element directly.
  • When working with TISAB reagent tablets or any chemical solutions, wear appropriate personal protective equipment, including gloves and eye protection.
  • Do not immerse the entire meter in liquid. The meter is designed for electrode immersion only.
  • Store the meter and reagents in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Çocukların erişemeyeceği yerde saklayınız.

3. Ürün Bittiview

The Extech FL700 Fluoride Meter is a compact and user-friendly device. Key components and features are outlined below:

Extech FL700 Fluoride Meter front view
Ön view of the Extech FL700 Fluoride Meter, showing the display and control buttons.
Hand holding Extech FL700 Fluoride Meter
The Extech FL700 Fluoride Meter held in hand, demonstrating its ergonomic design and clear display.
Close-up of Extech FL700 electrode tip
Kapatmak view of the fluoride electrode tip, which is integrated into the meter.
Extech FL700 Fluoride Meter in use with sample
The Extech FL700 Fluoride Meter being used to measure a liquid sample, showing the meter in its stand.
Extech FL700 Fluoride Meter and included accessories
All components included with the Extech FL700: the meter, sensor cap, TISAB reagent tablets, and neckstrap.

Temel Özellikler:

  • İki noktalı kalibrasyon: Ensures accurate measurements across the fluoride range.
  • Bellek depolama: Stores up to 25 labeled readings for later review.
  • Field or lab testing: Designed for versatile use in various environments.
  • Fast response: Provides fluoride measurements in less than 1 minute.
  • Large dual LCD: Simultaneously displays ppm (or mg/L) and temperature with a bargraph.

4. Kurulum

4.1. Pil Kurulumu

The Extech FL700 requires four 3V button batteries (included). To install or replace batteries:

  1. Twist off the battery compartment cap located at the top of the meter.
  2. Insert the four 3V button batteries, ensuring correct polarity as indicated inside the compartment.
  3. Replace the battery compartment cap and twist to secure.

4.2. İlk Hazırlık

Before first use or after prolonged storage, prepare the electrode:

  1. Remove the sensor cap from the electrode tip.
  2. Rinse the electrode tip with distilled water.
  3. Condition the electrode by immersing it in a fluoride standard solution (e.g., 1 ppm or 10 ppm) for approximately 15-30 minutes.

5. Kullanım Talimatları

5.1. Açma/Kapama

Basın AÇIK/KAPALI Sayaç açma veya kapatma düğmesi.

5.2. Calibration (Two-Point)

Accurate calibration is crucial for reliable measurements. The FL700 supports two-point calibration.

  1. Prepare two fluoride standard solutions, typically one low concentration (e.g., 1 ppm) and one high concentration (e.g., 10 ppm). Ensure the solutions are at a stable temperature.
  2. Elektrodu damıtılmış suyla durulayın ve kurulayın.
  3. Immerse the electrode into the first standard solution.
  4. Basılı tutun KAL/GERİ ÇAĞIRMA button until 'CAL' appears on the display. The meter will automatically recognize the standard and calibrate. Wait for the reading to stabilize.
  5. Rinse the electrode again and immerse it into the second standard solution.
  6. Basılı tutun KAL/GERİ ÇAĞIRMA button again. The meter will complete the two-point calibration.
  7. The meter is now calibrated and ready for measurement.

5.3. Ölçüm Alma

To measure fluoride concentration in a samptarih:

  1. Ölçüm cihazının kalibre edildiğinden emin olun.
  2. Elektrodu damıtılmış suyla durulayın ve kurulayın.
  3. Add TISAB (Total Ionic Strength Adjustment Buffer) reagent to your sample according to the reagent instructions. TISAB ensures that the fluoride ions are in a measurable form and minimizes interference.
  4. Immerse the electrode tip into the prepared sample solution. Ensure the tip is fully submerged.
  5. Allow the reading to stabilize on the display. The large dual LCD will show the fluoride concentration in ppm (or mg/L) and the temperature.
  6. Geçerli okumayı kaydetmek için, tuşuna basın. MOD/BEKLET düğmesine basın. Serbest bırakmak için tekrar basın.

5.4. Recalling Stored Readings

The meter can store up to 25 labeled readings.

  • To recall stored readings, press the KAL/GERİ ÇAĞIRMA button briefly when not in calibration mode.
  • Kullanın MOD/BEKLET button to scroll through the stored readings.

6. Bakım

6.1. Elektrot Temizliği

Regular cleaning of the electrode ensures optimal performance.

  • Her kullanımdan sonra elektrodu damıtılmış suyla iyice durulayın.
  • If the electrode becomes fouled or readings become erratic, clean it with a mild detergent solution or a specialized electrode cleaning solution. Rinse thoroughly with distilled water afterward.

6.2. Depolama

Proper storage extends the life of the electrode and meter.

  • Always replace the sensor cap on the electrode tip when not in use. The cap should contain a small amount of electrode storage solution or a fluoride standard solution to keep the membrane hydrated. Do not store dry.
  • Sayacı serin ve kuru bir yerde saklayın.

6.3. Pil Değişimi

Replace batteries when the low battery indicator appears on the display or when the meter fails to power on. Refer to Section 4.1 for battery installation instructions.

7. Sorun Giderme

SorunOlası NedenÇözüm
Sayaç açılmıyorBitmiş veya yanlış takılmış pillerPil kutuplarını kontrol edin; pilleri değiştirin.
Erratic or unstable readingsDirty electrode; air bubbles on electrode; insufficient sample volume; expired or incorrect TISAB.Clean electrode; gently tap electrode to remove bubbles; ensure electrode is fully submerged; check TISAB.
Meter fails to calibrateIncorrect standard solutions; dirty electrode; electrode not conditioned; damaged electrode.Verify standard concentrations; clean and condition electrode; replace electrode if damaged.
Düşük pil göstergesiPiller zayıf.Replace all four 3V button batteries.

8. Özellikler

ŞartnameDetay
ÜreticiEkstech
Parça NumarasıFL700
Ürün Ağırlığı7.7 ons
Ürün Boyutları6.8 x 1.6 x 1.4 inç
Ürün Model NumarasıFL700
Piller4 Lityum Metal pil gereklidir (dahil)
StilFL700
Ürün Paket Miktarı1
Parça Sayısı1
Dahil BileşenlerBirim, Talimat Kılavuzu
Pil Hücresi TipiAlkali
İlk Mevcut Tarih15 Temmuz 2008

9. Garanti Bilgileri

The Extech FL700 Fluoride Meter comes with a full one-year warranty on parts and workmanship, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modification.

10. Destek

For technical support, service, or additional information, please visit the official Extech webSitelerini ziyaret edebilir veya müşteri hizmetleri departmanıyla iletişime geçebilirsiniz. Ayrıca daha fazla bilgi ve ürüne şu adresten ulaşabilirsiniz: Extech Mağazası Amazon'da.

İlgili Belgeler - FL700

Önview EXTECH 45170 4'ü 1 Arada Nem, Sıcaklık, Hava Akışı ve Işık Ölçer Kullanım Kılavuzu
Nem, sıcaklık, hava akışı ve ışık ölçümlerini yapan 4'ü 1 arada ölçüm cihazı EXTECH 45170'in kullanım kılavuzu. Kullanım kılavuzu, teknik özellikler ve garanti bilgilerini içerir.
Önview UV-AB Extech UV505 Radiazione Kullanım Kılavuzu
Kullanım kılavuzu UV-AB Extech UV505 Radyatör Filtresi için tamamlandı. Operasyonel işlemleri, özel teknikleri, güvenlik ve bakım bilgilerini içerir.
Önview Extech UV505 UV-AB Işık Ölçer Kullanım Kılavuzu
Extech UV505 UV-AB Işık Ölçer kullanım kılavuzu. Bu belge, cihazın özellikleri, çalışması, teknik özellikleri, güvenlik önlemleri ve UVA ve UVB ışığını ölçmek için gereken bakım hakkında ayrıntılı bilgi sağlar.
Önview Extech EX810 1000 Amp Klemmeter IR Termometre Gebruikershandleiding ile tanıştı
Extech EX810 1000 ile ilgili ayrıntılar Amp Klemmetre ingebouwde IR termometreyle buluştu. Kapsama, ayaklık, arka duruş, kapasite, sıklık ve sıcaklık gibi yeni metinler için talimatlar, zengin bilgiler ve teknik özellikler.
Önview Extech EX650 Serisi Kullanım Kılavuzu: 600A Gerçek RMS Dijital Akım Ölçeramp Metreler
Extech EX650 Serisi 600A Gerçek RMS dijital akım koruma cihazları için kapsamlı kullanıcı kılavuzu.amp EX650 ve EX655 modellerini kapsayan ölçüm cihazları. Doğru AC/DC voltaj ölçümü için çalışma prensibi, özellikleri, güvenliği ve teknik özellikleri hakkında bilgi edinin.tage, akım, direnç ve daha birçok ölçüm.
Önview Extech DM100 Cep Multimetresi Kullanım Kılavuzu
Extech DM100 4000 Sayımlı Otomatik Menzilli Cep Multimetresi için güvenlik, teknik özellikler, kullanım, fonksiyonlar ve bakım konularını kapsayan kullanıcı kılavuzu.