Maxtec R217P42

Maxtec MAXO2+ Oksijen Analiz Cihazı Kullanım Kılavuzu

Model: R217P42

1. Giriş

The Maxtec MAXO2+ Oxygen Analyzer is a portable device designed for accurate measurement of oxygen concentration. It is commonly used in various applications, including checking oxygen levels in scuba tanks, medical oxygen sources, and other gas mixtures. This manual provides essential information for the safe and effective use of your analyzer.

2. Güvenlik Bilgileri

Please read and understand all safety warnings and instructions before operating the device. Failure to do so may result in injury or damage to the product.

3. Paket İçeriği

Paketinizde tüm öğelerin mevcut olduğundan emin olun:

4. Ürün Bittiview

The MAXO2+ Oxygen Analyzer features a compact design with an easy-to-read digital display. Key components include:

Maxtec MAXO2+ Oxygen Analyzer showing 20.9% oxygen reading

Figure 1: Maxtec MAXO2+ Oxygen Analyzer with digital display showing 20.9% oxygen.

5. Kurulum ve İlk Kullanım

Follow these steps to prepare your analyzer for use:

  1. Pilleri Takın:
    • Locate the battery compartment cover on the back of the analyzer.
    • Use the provided small screwdriver to unscrew the retaining screws and remove the cover.
    • 3 adet AAA pili, kutuplarının doğru olduğundan emin olarak takın.
    • Kapağı değiştirin ve vidalarla sabitleyin.
  2. Install Oxygen Sensor:
    • The oxygen sensor is typically packaged separately to prevent premature exposure to air.
    • Unscrew the cap from the sensor port at the bottom of the analyzer.
    • Carefully remove the new oxygen sensor from its protective packaging.
    • Screw the sensor into the sensor port until it is finger-tight. Do not overtighten.
  3. Attach T-Piece Adapter:
    • The T-piece adapter allows for sampling from various gas sources.
    • Firmly push the T-piece adapter onto the top port of the analyzer.
    • If using tubing for inline sampling, connect one end of the tubing to one side of the T-piece.

Video 1: Demonstration of the Oxygen Analyzer's features, T-piece attachment, and sampLing süreci.

6. Kullanım Talimatları

To ensure accurate readings, proper operation and calibration are crucial.

  1. Güç Açık: Press the power button (usually marked with an 'I' or power symbol) to turn on the analyzer. The display will illuminate and show the current oxygen concentration.
  2. Calibration (Fresh Air):
    • Before each use, or if the device has been stored for a long time, calibrate the analyzer in fresh air (20.9% oxygen).
    • Ensure the T-piece is open to ambient air, or no gas source is connected.
    • Press and hold the calibration button (often marked with a 'CAL' or similar symbol) until the display shows "20.9%" or indicates successful calibration. Some models may have a one-button calibration feature.
    • Okumanın sabitlenmesini bekleyin.
  3. Ölçüm Alma:
    • Connect the T-piece adapter to your oxygen source (e.g., scuba tank valve, oxygen concentrator outlet). For scuba tanks, ensure a snug fit with the rubberized dome.
    • Slowly open the valve of the oxygen source to allow gas to flow through the T-piece and over the sensor.
    • Observe the digital display for the oxygen concentration reading. Allow a few seconds for the reading to stabilize.
    • Once the reading is stable, note the measurement.
    • Close the oxygen source valve and disconnect the analyzer.
  4. Kapat: Ekran kapanana kadar güç düğmesini basılı tutun.

7. Bakım

Regular maintenance ensures the longevity and accuracy of your MAXO2+ Oxygen Analyzer.

8. Sorun Giderme

SorunOlası NedenÇözüm
Cihaz açılmıyor.Piller bitmiş veya yanlış takılmış.Pilleri doğru kutuplara yerleştirerek değiştirin.
Hatalı veya düzensiz okumalar.Sensor needs calibration, old sensor, or sensor not properly installed.Perform fresh air calibration. If problem persists, replace the sensor. Ensure sensor is tightly screwed in.
Calibration fails.Sensor is at the end of its life, or calibration environment is not fresh air.Replace the oxygen sensor. Ensure calibration is performed in an area with known 20.9% oxygen.
No gas flow through T-piece.Obstruction in tubing or T-piece, or gas source not properly connected/opened.Check tubing for kinks. Ensure T-piece is clear. Verify gas source connection and valve.

9. Özellikler

10. Garanti ve Destek

Garanti bilgileri veya teknik destek için lütfen üreticinin resmi web sitesine başvurun. webSite üzerinden veya müşteri hizmetleri departmanıyla iletişime geçin. Satın alma belgenizi saklayın.

İlgili Belgeler - R217P42

Önview MaxO2+ Endüstriyel Oksijen Analizörü: Kullanım Talimatları
Bu belge, Maxtec MaxO2+ Endüstriyel Oksijen Analizörü için özelliklerini, çalışmasını, kalibrasyonunu, bakımını, sorun gidermeyi, teknik özelliklerini ve elektromanyetik uyumluluğunu kapsayan kapsamlı talimatlar sağlar.
Önview Maxtec MAXO2+ AE Endüstriyel Oksijen Analizörü: Kullanım Kılavuzu ve Talimatlar
Maxtec MAXO2+ AE Endüstriyel oksijen analizörü için kapsamlı kullanım kılavuzu ve talimatlar. Aşağıdaki sistemleri kapsar:view, işletme, kalibrasyon, bakım, özellikler ve sorun giderme.
Önview Maxtec MaxO2+ Oksijen Analiz Cihazı Kullanım Kılavuzu
Maxtec MaxO2+ Oksijen Analiz Cihazı için kapsamlı kullanım kılavuzu; tıbbi ortamlarda doğru oksijen ölçümü için çalıştırma, kalibrasyon, bakım, teknik özellikler ve sorun giderme konularını ayrıntılı olarak açıklamaktadır.
Önview MaxO2+A Oksijen Analizörü: Özellikler, Teknik Özellikler ve Aksesuarlar
Maxtec'in taşınabilir ve kullanıcı dostu oksijen analizörü MaxO2+A'yı keşfedin. Temel özellikleri, teknik özellikleri, parça numaraları ve çeşitli uygulamalarda doğru oksijen ölçümü için mevcut aksesuarları hakkında bilgi edinin.
Önview Maxtec Handi+ Endüstriyel Oksijen Analizörü: Kullanım Talimatları
Maxtec Handi+ endüstriyel oksijen analizörü için kurulum, çalıştırma, kalibrasyon, hata kodları, bakım ve teknik özellikleri kapsayan kapsamlı talimatlar. Hassas izleme için MAX-250 oksijen sensörüne sahiptir.
Önview MaxBlend 2 Düşük Akışlı / Yüksek Akışlı Oksijen Karıştırıcı - Kullanım Talimatları
This document provides comprehensive instructions for use, safety guidelines, specifications, and troubleshooting for the Maxtec MaxBlend 2 Low Flow / High Flow Oxygen Blender, a medical device for air/oxygen gas mixing and monitoring.